Set temperature range in °C | min. 5°C above ambient up to +80°C |
---|---|
Setting accuracy temperature | 0.1°C |
Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE | adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
Sterilisation | fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load |
Convection | natural convection |
---|---|
Fresh air admixture | adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
Vent | vent connection with restrictor flap |
Documentation | programme stored in case of power failure |
---|---|
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
---|---|
Temperature control | overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY | additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system | for fault analysis |
Alarm | visual and acoustic |
Works calibration certificate | incl. works calibration certificate for +37°C |
---|---|
Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
Door | inner glass door |
Internals | 2 stainless steel grids |
Interior | easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
---|---|
Volume | 256 l |
Dimensions W x H x D in mm | w(A) x h(B) x d(C): 640 x 800 x 500 mm |
Max. number of grids/shelves | 9 |
Max. loading of chamber | 300 kg |
Max. loading per grid/shelf | 20 kg |
Dimensions | w(D) x h(E) x d(F): 824 x 1183 x 684 mm |
---|---|
Housing | rear zinc-plated steel |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1700 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 900 W |
the appliances must be transported upright | |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | B x H x T: 930 x 1380 x 930 mm |
Net weight | approx. 110 kg |
Gross weight carton | approx. 161 kg |